Repository logo
Communities & Collections
All of DSpace
  • English
  • العربية
  • বাংলা
  • Català
  • Čeština
  • Deutsch
  • Ελληνικά
  • Español
  • Suomi
  • Français
  • Gàidhlig
  • हिंदी
  • Magyar
  • Italiano
  • Қазақ
  • Latviešu
  • Nederlands
  • Polski
  • Português
  • Português do Brasil
  • Srpski (lat)
  • Српски
  • Svenska
  • Türkçe
  • Yкраї́нська
  • Tiếng Việt
Log In
New user? Click here to register.Have you forgotten your password?
  1. Home
  2. Browse by Author

Browsing by Author "Harshanee, K.G.A.T."

Filter results by typing the first few letters
Now showing 1 - 2 of 2
  • Results Per Page
  • Sort Options
  • Thumbnail Image
    Item
    Comparative analysis of alkaline phosphatase with two assays using different buffers; diethanolamine (dea) and 2-amino-2-methyl-1-propanol (amp): establishing correlation factors for diagnostic consistency
    (College of Chemical Pathologists of Sri Lanka, 2024) Jayasekara, D.; Fernando, K.; Kulasinghe, M.; Silva, P.; Madurangi, D.W.D.D.; De Silva, D.D.S.; Harshanee, K.G.A.T.; Bandara, S.R.R.; Dayanath, B.K.T.P.
    INTRODUCTION AND OBJECTIVES Alkaline phosphatase (ALP) serves as a pivotal biomarker for bone and liver diseases, employing assays utilizing either 2-amino-2-methyl-1-propanol (AMP) (IFCC recommended) or diethanolamine (DEA) buffers, with the latter consistently yielding higher values. This study aimed to develop a correlation factor for ALP reagents using DEA buffer from supplier X, in comparison to routine automated ALP assay at the central laboratory using AMP.METHODS Twenty-five serum samples were analyzed in the central laboratory assay using AMP buffer in a fully automated analyzer with dedicated reagents and the test assay using DEA buffer on a semi-automated biochemistry analyzer within two hours of receipt. Both assays employed the same biochemical reaction, differing only in buffer composition. The linearity ranges for the test assay with DEA buffer and the routine assay with AMP buffer were determined as 1600 U/L and 800 U/L, respectively. RESULTS Patient samples exhibited ALP levels ranging from 0 to 339 U/L by routine assay. The correlation graph demonstrated a satisfactory R2>0.75, indicating adequate number of sample inclusion and quality. A correction factor of 1.2 was calculated for the ALP assay utilizing DEA, compared to the AMP-based assay, employing simple linear regression analysis.CONCLUSIONS According to the sample availability, only ALP levels up to 339 U/L by AMP-based assay were included. Therefore, the correction factor of 1.2 is applicable only up to an ALP level of 400 U/L with the DEA-based assay, necessitating dilution of samples with higher values for the correlation factor’s application. This study indicates a correction factor of 1.2, which is deviated from factors close to 2, observed in literature because of reagents being from different manufacturers and running two assays on two different platforms (automated/ semiautomated). It is important to derive a factor for an ALP assay with DEA buffer to make the results comparable to IFCC recommended AMP buffer used ALP assay.
  • Thumbnail Image
    Item
    Evaluation of point-of-care testing (poct) devices for cardiac troponin i in screening patients with myocardial infarction
    (College of Chemical Pathologists of Sri Lanka, 2024) Fernando, K.; Jayasekara, D.; Kulasinghe, M.; Silva, P.; Harshanee, K.G.A.T.; Bandara, S.R.R.; Dayanath, B.K.T.P.
    INTRODUCTION AND OBJECTIVES Cardiac troponin I (cTnI) is a crucial biomarker for diagnosing myocardial infarction (MI). However, many remote hospitals lack access to cTnI assessment facilities. This study investigates the feasibility of using Point-of Care Testing (POCT) devices to triage MI patients in such settings, facilitating their transfer to tertiary care hospitals. Assessing the quality of POCT devices is essential for this purpose. This report outlines the assessment methodology of two POCT devices and presents the obtained results.METHODS Two POCT devices, labelled X and Y, for measuring cTnI were compared against the Ortho-Vitros 3600, serving as the reference method, with optimum internal and external quality control measures. Basic specifications of the POCT devices and the comparator were obtained from their respective kit inserts. Routine patient samples were analyzed in singlicate using POCT devices and the reference method. Linear regression analysis was conducted, and correlation graphs were generated. Within-run precision was evaluated using a patient sample and imprecision (CV) was calculated for the POCT devices.RESULTS The linearity ranges for cTnI measurement with POCT devices X and Y were 0.01–15 ng/mL and 0.03–30 ng/ mL, respectively, with decision-making cutoff values for diagnosing MI established at 0.04 ng/mL and 0.5 ng/ mL, respectively. Regression analysis demonstrated acceptable linearity for both POCT devices, with correlation of R2=0.7388 for device X and R2=0.8881 for device Y. However, higher imprecision was observed for both analyzers, with a CV of 26% for device X and 20% for device Y at decision-making cutoff levels.CONCLUSIONS This study highlights major challenges associated with the implementing POCT devices for cTnI assays in triage settings aimed at diagnosing MI in acute care. Specifically, the adoption of higher cutoff levels in the POCT devices, not aligned with the recommended 99th percentile for the healthy population, and the poor precision observed at these cutoff values emerge as significant hurdles.

DSpace software copyright © 2002-2025 LYRASIS

  • Privacy policy
  • End User Agreement
  • Send Feedback
Repository logo COAR Notify