Does Sanders-Retzlaff-Kraff (SRK-2) formula suit a Sri Lankan population in phacoemulsification cataract surgery of medium axial length?

No Thumbnail Available

Date

2012

Journal Title

Journal ISSN

Volume Title

Publisher

Sri Lanka Medical Association

Abstract

INTRODUCTION: After introduction of phacoemulsification with small incision techniques, intraocular lens (IOL) power calculation predictionhas become a crucial step for good refractive outcomes. SRK-2 formula is widely used for that purpose in many locations. AIMS: To evaluate the predictive accuracy of post operative refraction using SRK-2 formula in phacoemulsification of eyes with 22-26mm axial length. METHODS: Patients with senile cataracts undergoing phacoemulsification and IOL during 2011 in Gampaha district by a single surgeon were selected randomly. Single examiner measured simulated manual keratometry preoperatively, seven days postoperatively. Uncorrected visual acuity, slit lamp examination of anterior segment and best spectacle-corrected visual acuity (BSCVA) were measured. The implanted IOL power was used to calculate the predicted postoperative refractive error using time SRK-2 formula. RESULTS: Of 274 patients (56% females, mean (SD) age 65.3(10) years), the right eye was operated in 142(51.8%). On postoperative seventh day, 79% had VA of 6/6 and 14% to 6/9, 4% to 6/12 to 6/36 in the operated eye. Mean (SD) of IOL power used 21.50 (1.90)D. The predicted refractive error with SRK-2 was -0.3(0.145)0 and the achieved refractive error was -0.22(0.732)0. Difference between predicted and achieved refractive error presented a slight hyperopic shift (mean(SD) 0.054(0.397)0}. There was a negative Pearson correlation (-0.126) between the predicted refractive error and achieved refractive error (p=0.04). Predictive error was less than 0.5 in 80.2%, <0.75 in 88.6% and <1.0 in 96.7%. CONCLUSIONS: SRK-2 formula is a good option to predict the refractive error after cataract extraction by phacoemulsification in eyes with medium axial length.

Description

Poster Presentation Abstract (PP 140), 125th Anniversary Scientific Medical Congress, Sri Lanka Medical Association, June 2012 Colombo, Sri Lanka

Keywords

(SRK-2)

Citation

Sri Lanka Medical Association, 125th International Medical Congress. 2012;57 Suppliment1: 138

Endorsement

Review

Supplemented By

Referenced By